Miniature Probes

Miniature Probes

Potentiometric oxygen probes developed for the exhaust gas measurements in motor vehicles (λ-probes) are frequently used in industrial equipment. Due to their construction, these probes have some disadvantages: because of the uncontrolled heating (steady working voltage) and ceramic coating of the electrodes (which increase the diffusion of special gas components) significant measuring errors can result.

The ZIROX miniature probes MS27 and SS27 with controlled heater and integrated primary electronics features high accuracy at comparatively lower costs. The main part is the approved ZIROX potentiometric zirconia-based solid electrolyte cell (drift-, calibration- and maintenance-free). The use of high-quality components and materials guarantees high long-term stability.

The SS27 was developed for trace measurements under oxidizing conditions. The MS27 was constructed for reducing conditions or oxygen concentration of more than 100 ppm. Both probe types only differ in the type of sensor element.  

 
Oxygen Measurement Applications
  • Oxygen measurements in protective gases
  • Exhaust gas measurements in large-scale engines
  • Control optimization in small combustion plants
  • Furnace gas measurements in heat treatment equipment
  • Optimization of incineration plants
  • Monitoring of industrial processes under protective or inert gas atmosphere
Technical Data Miniature Probe SS27/MS27
Installation length 65 mm (incl. M18x1.5), 75 mm (DN40KF)
Diameter 10 mm
Weight 0.5 kg
Dimension clamp head 64 x 58 x 36 mm
Protection degree IP42, others on request
Power supply 24 V DC
Power consumption 17 W
Warm-up approx. 5 min
Sensor voltage (raw signal) 0…200 mV (0…300 mV on request)
Range (SS27) 15 ppm…20.6 Vol.-% O2 (0…200 mV) 0.13ppm…20.6 Vol.-% (0…300 mV)
Accuracy < 5 % rel. error
Measuring gas temperature max. 300 °C
Probe working temperature 700 °C
Surrounding conditions probe head max. 50 °C, rel. humidity < 80 %
Offset approx. -2…-8 mV (compensable)
Output signal 4-20 mA (0-5 V, 0-10 V on request)
Plug Assignments
(1) Analog output + IA
(2) Analog output GND IA
(3) Power supply GND VB
(4) Power supply + 24 V VB
(5) Relay RDY
Calculation of Oxygen Concentration
The oxygen partial pressure is calculated by the NERNST equation. ϕ(O₂) = 20.64 * e^(-46.42 U/T)
  • ϕ(O₂): oxygen concentration in Vol.-%
  • U: cell voltage in mV
  • T: cell temperature in K (700 °C = 973.15 K)
  • 20.64: oxygen concentration in ambient air at 50% relative humidity
Translation of the Output Signal (Range 0…200mV)
Current output 4-20 mA 1 mA ≙ 12.5 mV cell voltage
Voltage output 0-5 V 1 V ≙ 40 mV cell voltage
Voltage output 0-10 V 1 V ≙ 20 mV cell voltage
Example for 0…200 mV range:
At a current of 5.5 mA ((5.5-4)mA × 12.5 mV = 18.75 mV) an oxygen concentration ϕ(O₂) = 8.43 Vol.-% results.
Download PDF Download ⬇️