H2-Measuring Device TCS

General Information
Capsule Pressure Gauge
SWM2000

Oxygen Hydrogen Measuring Device. Main parts are a zirconia oxygen sensor and a TCD for the hydrogen detection. Additionally, further thermodynamic values (or data) can be calculated on the basis of the oxygen/hydrogen measurement. We provide more information on request. Please contact us by e-mail.

SGM5-NGW

This device serves the determination of the Non-Equilibrium-Oxygen (NGW-Sauerstoff in German) in reducing atmosphere. It is mainly used in applications where the equilibrium oxygen cannot be used as target value. As for the devices before, a detailed consulting is necessary for correct application. Please contact us per e-mail.

EC-Module

Inexpensive oxygen measuring device based on aqueous electrolyte systems. It is mainly used for the oxygen detection in the percent range. For further information contact us by e-mail please.

Hydrogen Sensor TCS

The hydrogen sensor TCS (thermal conductivity sensor) is a compact measuring device, which is directly mounted on the measuring chamber by vacuum components. The sensor is pre-thermostatted by its own electronics, whereby a very high stable sensor signal is ensured. Originally developed for nitriding equipment of contract heat treatment workshops, the sensor is applicable for special measuring problems (e.g. low pressure application) due to a sophisticated calibration method.

H2-Measuring Device TCS
Applications

In many technological processes, hydrogen plays a big role in achieving the desired product properties. By means of TCS, the hydrogen concentrations can be analyzed and provided to the process control via standard interfaces.

Due to the increasing adoption of quality management systems (e.g., ISO 9001), a constant monitoring and documentation of measured values is necessary.

With the TCS, the user can monitor hydrogen measurement efficiently and continuously.

Technical Data H2-Measuring Device TCS
Range 0...100 Vol.-%
Accuracy rel. error < 5 %
Response time t90 (sensor) approx. 180 s (depends on gas distribution at measuring chamber)
Sensor temperature 50 °C (pre-thermostatted, electronically controlled)
Dimensions (WxHxD) 80 mm × 125 mm × 83 mm
Signal output analog output 0/4-20 mA (scaling by software) and serial interface RS232 (0-5/10 V on request)
Power supply 24 V DC +/- 20 %, approx. 20 W
Display none, operation via software
Protection degree IP 65
Ambient temperature 10...50 °C, rel. humidity 0...95 % at 20 °C
Storage temperature -20...60 °C, rel. humidity 0...95 % at 20 °C
Gas connection via DIN40KF (others on request)
Gas supply via diffusion (Gas distribution at measuring chamber)
Ambient pressure 950...1100 hPa (Low pressure applications need special calibration)
Note: Measuring value depends on pressure
Warm-up 10 min.
Calibration First calibration (characteristic line determination) by manufacturer, a 2-point calibration approx. every 3 months (process pressure must be known and provided to manufacturer) by user
Mounting No requirements
Cross-sensitivity Gases with high thermal conductivity falsify the measuring result

Download PDF Download ⬇️

Note: You are not sure which device is required for your application?
Please complete our questionnaire and send it to info@zirox.de Email Icon .