Vacuum-Resistant Oxygen Probe for PVD or Magnetron Sputtering

General Information

Main part of the ZIROX vacuum probe is a potentiometric zirconia solid electrolyte cell. Due to various oxygen partial pressures (inside and outside the vacuum recipient) a voltage occurs between measuring and reference electrode. This voltage is proportional to the logarithm of the oxygen partial pressures. From the cell voltage the oxygen partial pressure at measuring electrode can be calculated according to the NERNST equation. The connection head contains the electronics for heater control and signal processing and the membrane pump for the reference gas supply.

ZIROX Vacuum Probe

The ZIROX vacuum probe is used to determine oxygen partial pressures in low-pressure equipment. Typical applications include vacuum processes (PVD, magnetron sputtering, and other plasma processes), surface treatment applications, and material-scientific investigations.

Note:

In contrast to the ZIROX normal pressure probes, the vacuum probe is not calibration-free. The reasons are the special sensor heater (inside the probe tube) and several conditions at the measuring electrode (energy removal depends on gas pressure and thermal conductivity). Therefore, isothermal electrodes, required for the correct use of the NERNST equation, cannot be realized. For process control, however, the use of empirically determined values is sufficient in most cases. For the measurement of oxygen partial pressures, the user must calibrate the probe (test measurements).



ZIROX Vacuum Probe
Technical Data Vacuum Probe

Inbuilt length140...300 mm
Diameter6 mm, with protection tube 16 mm
Mass1.5 kg
Dimension clamp head150 x 63 x 35 mm
Protection degreeIP54
FlangeKF40, others on request
Max. temperature at flange60 °C
Power supply24 V DC +/- 10 %
Current consumption1.2 A
Heating power30 W
Pressure range1.5 x 10^5...1 x 10^-5 Pa
Sensor voltage range0...400 mV or 0...1200 mV (range changing on circuit board)
Range (oxygen partial pressure)1 x 10^5...1 x 10^-21 Pa
Output signal0-20 mA
Accuracy< 5 % rel. error
Sensor working temperature700 °C (electronically controlled)
Offset (with protection sleeve)-15...-20 mV
Helium leak rate< 10^-8 mbar l/s
Gas flowmax. 10 m/s
Reference gasSurrounding air
Reference gas flow5...10 l/h (by internal pump)
Plug Assignment
1 + 24 V
2 GND
3 + Iout 20 mA corresponds to 400 or
1200 mV (as adjusted)
4 - Iout
5 Ready contact (potential free) 60V/1A DC or
125V/1A AC
6
Download PDF Download ⬇️

Note: You are not sure which device is required for your application?
Please complete our questionnaire and send it to info@zirox.de Email Icon .